Skip to main content

Compensation in Flow Cytometry (Fluorescence Spill Over Correction).

                     The term "Compensation" in flow cytometers refers to the process of correcting fluorescence spill over. The compensation matrix ensures that the fluorescence emitted is detected in a particular detector which is from the fluorescence being measured. As we all know that most of the fluorescence molecules have broad emission range. if more than one one fluorescence is selected, their emission spectra may overlap. To correct this spectral overlap in cytometry data, a mathematical matrix called compensation is used.
The Intention of this article is to teach you the compensation from basic theory to practical.
This example for FITC and PE fluorescence molecules shows that, Some of the FITC fluorescence is being detected in PE Detector. and Some of the PE fluorescence is being detected in FITC Detectors. Task here is to make sure that in cytometry data, FITC is in FITC detector only and PE is in PE Detector only.

(FITC + PE Overlap) - (PE Overlap) = accurate FITC
(PE + FITC Overlap) - (FITC Overlap) = accurate PE
Source: BD Spectrum Viewer



Experiment: Analysis of two-color flow cytometry data in BD Accuri C6 Flow Cytometer. FITC and PE. And applying the Compensation.
Sample Tubes Required:
                  1) Unstained Cells
                  2) Cells with FITC
                  3) Cells with PE
                  4) Cells with both FITC and PE

Flow Cytometer: BD Accuri C6


1) Data Plots: SSC vs FSC, FITC vs PE. (If you want, you can draw another plot to identify the singlets, that is FSC-Height vs FSC-Area). For all tubes, make sure you record sufficient and equal number of cells. It would be better if you record 20,000 events for each tube. 

2) Load the tube with Unstained Cells and Record the data
3) Load the tube FITC cells and record the data 
4) Load the tube with PE Cells and record the data

Analyze the fluorescence data for FITC and PE. FITC is falling in PE Detector and some of the PE cells very few are falling FITC Detector. Now, Compensation must be applied.. Compensation values for these two fluorescence in Accuri C6 are shown in the below compensation matrix.

(FL1 is for FITC and FL2 is for PE). After applying these values spill over compensation values. You will see the data like shown below.
After compensation, FITC is in FITC detector and PE is PE Detector. 

5) Finally load the tube which has both FITC and PE Cells and record the data
    (Here there are four possibilities, 1: cell has only FITC, 2: cell has only PE and 3: Cell has both FITC and PE, 4: Cell has no fluorescence.. These possibilities must be considered when analyzing more than one fluorescence in flow cytometry). 
Since, compensation has been applied above for FITC and PE fluorescence, the sample tube which has both FITC and PE cells will show data like shown below.

That is it.. The above shown is for simple 2 color flow cytometry data analysis..
In today's flow cytometers, you can analyze up to 15-20 colors at a time. which is typically called  a multi-color flow cytometry..

For multi-color flow cytometry, you need to have good understanding of the instrument sensitivity, performance, understanding fluorescence excitation and emission, antibody titration, fluorochrome choice, required controls and many more parameters should be considered. 


Comments

Post a Comment

Popular posts from this blog

Nanodrop

Thermofisher, Nanodrop 1000    Nanodrop is a spectrophotometer which is designed to measure the concentration of nucleic acids at 260nm and Proteins at 280nm present in a very small minute volume. It is also capable of measuring the concentrations of dsDNA, RNA, ssRNA and purified protein. Required sample size is usually 1µL. Nanodrop instruments are usually full spectrum spectrophotometer, i.e., Ultra violet and visible range wavelengths (250nm to 750nm).  Nanodrop is commonly used for the quantification of the DNA concentration and RNA concentration in a given sample. However there are many other applications which can be performed on this instrument.  It works on the principle of Spectrophotometer or Sample Retention System....  Principle of Nanodrop: "The Nanodrop works on the principle of sample retention system.. When the small volume of sample is introduced, the arm and pedestal will work together and make the sample in the form of colum...

Statistics in Flow Cytometry Data and "MFI" values

                                 The speed of the flow cytometry offers wide range of data points and data plots. Due to its sensitivity and versatility, it has been used extensively. Flow cytometer can detect up to 1000s of cells per second.. But, it should be noted that the flow cytometry also involves the statistics, its significance, calculating the fluorescence intensity..                Statistics in flow cytometry involves, total number of data points acquired, percentage population, most importantly mean and median fluorescence intensity, and others. Most of us get confused about the term "MFI", whether it means mean fluorescence intensity or median fluorescence intensity.. Truth is, both.. It means it can be described as either mean or median fluorescence intensity. Some researchers use Mean for MFI and some use Median for MFI values.  (Keep...